Technische Daten
Oberflächenkraft: 1.0 ± 0.05N
Prüfspannung: 100 bis 600 V
Maximaler Prüfstrom:: 3a
Abstand zwischen zwei Elektroden:: 4,0 mm
Abwurfvorrichtung: Ablasszeitintervall kann beliebig eingestellt werden
Volumen der Prüfkammer: 0.5m3, Tiefe 60x Breite 95x Höhe 90cm
Externe Dimensionen: Tiefe x Breite x Höhe: 61x 120x 105cm
Außenkartonmaterial: Spiegel-Edelstahl
Hervorheben:

leakage trace testing machine for rubber

,

rubber testing equipment with warranty

,

leakage detection machine for materials

Grundinformation
Herkunftsort: China
Markenname: CHENGFENG
Zertifizierung: ISO,CE
Modellnummer: CF8352
Zahlung und Versand AGB
Verpackung Informationen: Holzkisten
Lieferzeit: 10 Arbeitstage
Zahlungsbedingungen: T/T, Cash.L/C, D/A, D/P, Western Union, MoneyGram
Versorgungsmaterial-Fähigkeit: 5 Einheiten pro Monat
Produkt-Beschreibung

Basic principle: Under the influence of humid and impurity environments on electrical products, insulation leakage may occur between charged parts of different polarities or between charged parts and grounded metal. The generated arc can cause breakdown and short circuit to the electrical appliance, material corrosion due to discharge, and even lead to fire.

The leakage trace test is a destructive test conducted on insulation materials to simulate the above situation. The experiment is to apply a certain voltage and drop a specified volume of contaminated liquid (0.1% NH 4 CL) at a fixed height (35mm) for 30 seconds on the surface of a solid insulation material, between platinum electrodes of a specified size(2mm×5mm), to evaluate the tolerance of the solid insulation material surface under the combined action of an electric field and a contaminated medium, and to determine its relative tracking index (CT1) and resistance to tracking index (PT1).

Compliant with standard: Meets the requirements of GB/T4207 testing standard.

 

Electrode: Two rectangular platinum electrodes with a cross-sectional area of 2mm×5mm. One end edge has a 30°slope.
Surface force: 1.0 ± 0.05N
Test voltage: 100-600V
Maximum test current: 3A
Distance between two electrodes: 4.0mm
Dropping device: Dropping time interval can be set arbitrarily
Test chamber volume: 0.5M³,depth 60x width95x height90cm
External dimensions: depth x width x height: 61x 120x 105cm
Outer box material: mirror stainless steel.