Heat Distortion And Vicat Softening Point Temperature Tester (Micro-Controlled) Mainly Used For The Determination Of Heat Deformation And Vicat Softening Point Temperature Of Polymer Materials

1 Set
MOQ
US$2000-$30000
Price
Heat Distortion And Vicat Softening Point Temperature Tester (Micro-Controlled) Mainly Used For The Determination Of Heat Deformation And Vicat Softening Point Temperature Of Polymer Materials
Features Gallery Product Description Chat Now
Features
Specifications
Test Range:: Room Temperature To 300℃
Heater Power:: 4KW;
Vicat Deflection:: 1mm.
Deformation Resolution Error:: 0.005mm
Heating Rate:: A=5℃/6min, B=12℃/6min;
Maximum Temperature Measurement Error:: ±0.5℃
Maximum Temperature Control Error:: ±12℃/6 Minutes
Main Power Supply:: 220V+10%, 15A, 50HZ
Highlight:

micro-controlled heat distortion tester

,

polymer material vicat softening tester

,

heat deformation temperature testing equipment

Basic Infomation
Place of Origin: China
Brand Name: CHENGFENG
Certification: ISO,CE
Model Number: CF8534
Payment & Shipping Terms
Packaging Details: Wooden cases
Delivery Time: 10 Work Days
Payment Terms: T/T, Cash.L/C, D/A, D/P, Western Union, MoneyGram
Supply Ability: 5 units per month
Product Description

Scope of application
Applicable scope:
It is mainly used for the determination of heat deformation and Vicat softening point temperature of polymer materials (such as plastics, rubber, nylon, electrical insulating materials, etc.). It is applicable to colleges and universities, research institutions and product quality supervision and inspection units.
Compliant with standards
Complies with GB/T 8802-2001, ISO 2570 hard polyethylene (PVC), GB/T 1633, GB/T
Determination of Vicat softening temperature (VST) - GB/T 1633-2000(idt ISO 306:1994
GB/T 1634.1-2004(idt ISO75-2:2003) Determination of deformation temperature of plastics under load (General test Method)
GB/T 1634.2-2004 Determination of deformation temperature under load of plastics (plastics, hard rubbers and long fiber reinforced composites)
GB/T 1634.3-2004 Determination of deformation temperature of plastics under load (High Strength thermosetting laminated materials)
Determination of Vicat softening temperature of thermoplastic pipes and fittings - GB/T 8802-2001(eqv ISO 2507:1995
Main performance indicators
Test range: Room temperature to 300℃
Number of sample racks: 2, 3, 6;
Heater power: 4KW;
Vicat deflection: 1mm.
Deformation resolution error: 0.005mm
Heating rate: A=5℃/6min, B=12℃/6min;
Maximum temperature measurement error: ±0.5℃
Maximum temperature control error: ±12℃/6 minutes (thermal deformation test) ±5℃/6 minutes (Vicat test)
Weight of 1# weight (i.e., load rod, standard indenter, sensor probe, tray, screw) : 68g±1g
Heating medium: Silicone oil (less than 200 centistokes, flash point above 300℃)
Cooling method: Natural cooling or water cooling (C-type machine)
Overall dimensions: 520×420×1100mm;
Main power supply: 220V+10%, 15A, 50HZ;
A set of test application software for the Chinese version operating system of the heat distortion and Vicat softening point temperature tester

Recommended Products
Get in touch with us
Contact Person : Mia Feng
Tel : +86-13128029450
Characters Remaining(20/3000)